|
Using
the MAS (Multiwell AutoSampler) on the FACSCalibur
Table of contents
Introduction
Outline of an experiment using the MAS
Materials
Fluidics system for the MAS.
Instrument setup using CellQuest.
Using the MPM software
Summary of differences between MPM and WorkList Manager.
Introduction
With the MAS, you can acquire your FACS data from cells in 96 well
plate format. Samples can be in traditional round
bottom or V-bottom plates, or in 1- or 2-ml deep well plates
(although only
the
specific plates listed in Materials
and Equipment can be used). This permits
you to perform staining procedures in 96 well plates,
including whole blood protocols, saving you significant time.
This document describes how to use the MAS, including
materials, instrument setup, and using the MPM
software.
Outline of an experiment using the MAS
1. Stain you cells in a plate compatible with the
MAS. Include compensation samples. Make sure that
the volume
of the buffer
used for the final
resuspension of the cells does not vary from well-to-well.
2. Transfer your compensation samples to FACS tubes
(12 x 75 mm, from BD).
3. Check the volume of fluid in the sheath cubtainer
and the waste cubtainers. If appropriate, either
replace the
sheath
cubtainer or replace or empty the waste cubtainer.
4. Turn on the components of the cytometer in the
following order: (a) the FACSFlow supply, (b) the
MultiWell Autosampler
power,
(c) the cytometer, and (d) the computer.
5. Take the outer metal tube of the droplet containment
module (DCM) off the sample injection port (SIP).
6. Launch CellQuest and use it to obtain your compensation
and PMT settings. Save the Instrument Settings in
your own folder in the FACS
Users folder. Unlike Worklist Manager, the settings file does not
have to be stored in the "BD Files:Instrument Settings"
folder.
7. In CellQuest, construct a CellQuest Template
Document that you will use in the MPM software.
The CellQuest
Template will
contain
typical
acquisition plots. You must set Acquisition Parameters
to use "Events
or Time", and the time must be the lesser of 250 seconds or
the volume of your sample in microliters. The data in the Parameters
Description
box will not be used by the MPM software (nor are Panels used, in
contrast to WorkList Manager). Save the CellQuest
Template document
in you own
folder in the FACS Users folder.
8. Install the Cytometer Interface Unit (CIU) on
the SIP.
9. Launch the MPM software.
10. Set the cytometer on "Run" and "High". Run
the Prime System command from the MPM Autosampler menu two times.
11. Set up MPM parameters.
12. Run MPM.
13. When you are finished running your samples,
with the CIU in place run the Prime from the Autosampler
menu 3
times.
14. Remove the CIU from the SIP, replace the outer
tube of the DCM, and manually wash the cytometer
system for
5 minutes
with
10% bleach,
followed by 5 minutes with 10% water.
15. Transfer your data to an appropriate place.
16. Turn everything off.
Materials and Equipment
Plates
Only plates on the list specified in the document Plate_Definition_Guide.pdf may be used with the MAS. The MAS contains a sample probe that
is programmed to descend into the wells of your sample plate. The
depth that the probe descends is set by parameters that are associated
with a particular plate. The plates that we are most likely to
use are listed in the table below. Make sure that you have some
of these on hand.
351190
|
351190
|
96-well, storage plate, 340 µl volume |
PP
|
U
|
12777-030
|
3365
|
355263
|
355263
|
96-well, storage |
PP
|
V
|
BD353263
|
3363
|
353910
|
353910
|
96-well, clear |
PS
|
U
|
62406-015
|
3370
|
353964
|
353964
|
96-well, deep well, 1 µl |
PP
|
Conical
|
62406-494
|
3959
|
353966
|
353966
|
96-well, deep well, 2 µl |
PP
|
Conical
|
62406-496
|
3961
|
353968
|
353968
|
96-well, deep well, 350 µl
|
PP
|
Conical
|
62406-510
|
-
|
Vortexer
When vortexing samples in deep well plates, it is necessary to
use a vortex mixer that does not have significant up-and-down action.
We prefer the Titer
Plate Shaker Lab-Line (VWR catalog #57019-600).
Multichannel pipettors.
Alternatively, samples can be mixed by pipetting up and down.
For the large volumes associated with work with whole blood, multichannel
pipetters are available from Matrix
Technologies that are programmable and that deliver up to 1250 µl, and
are capable of mixing. See the following table for features.
| Impact |
6004
|
Fixed
|
8
|
1250
|
1
|
$1099
|
| Impact² |
2004
|
Fixed
|
8
|
1250
|
6
|
$1319
|
| Equalizer Impact |
6034
|
Variable
|
8
|
1250
|
1
|
$1575
|
| Equalizer Impact² |
2034
|
Variable
|
8
|
1250
|
6
|
$1794
|
The catalog
numbers for compatible pipette tips are found at the
Matrix web site.
Bottle-top dispensers
If an 8-channel pipetter that delivers ≥ 1 ml is not available,
it is possible to use a bottle-top dispenser. The dispenser must
be used to deliver the liquid with sufficient force to achieve
mixing, but gently enough to prevent contamination of samples between
wells. The Bottle-Top Dispenser from Barnstead[JDA2], or equivalent,
may be used.
Fluidics system of the MAS
The following is from the BD pdf document BD
Multiwell AutoSampler User's Guide.
The FACSFlow Supply System is an automated sheath
and waste fluid control system designed for use
with the
BD Multiwell
AutoSampler.
The FACSFlow Supply System includes:
- FACSFlow Supply System unit with two removable
bottom plates
- modified four-liter sheath tank for the cytometer
- modified four-liter waste tank for the cytometer
- probe sets and tubing with color-coded connectors
- probe holders
- sheath and air filters
A complete description of the FACSFlow system can
be found in the PDF document BD
Multiwell AutoSampler User's Guide,
present
in
the printed
manual for the MAS, or available on the Mordred
or FACS Lab computers.
The Cytometer Interface
Unit (CIU) is installed in place of a FACS tube when performing
acquisition from plates. Otherwise, conventional 12 x 75 tubes
can be used. When not in use, the CIU is mounted on the FACS
Calibur via magnets found on the top or the side of the CIU.
Before installing the CIU, it is necessary to remove
the Droplet Containment
Module (DCM) shield from
the Sample
Injection Port
(SIP). Unscrew the
fitting that holds the DCM shield onto the SIP,
slide the DCM shield down off of the SIP, slide
the screw
fitting off of
the DCM shield,
slide it back onto the SIP, and rescrew it into
place. Set the DCM shield in a safe place, as you
must reinstall
it
when you
are finished
with the MAS.
You will not need to manipulate any of the fluidics
system behind the plate door of the MAS.
Setting Things up in CellQuest
Before running MPM, it is necessary to run CellQuest for the following
reasons:
- Obtaining instrument settings, which may be stored in your
own user directory.
- Setting up and saving a CellQuest template, which includes
appropriate acquisition dot plots, gates, and statistics panels.
This template can be saved anywhere as well, though I recommend
that you save it in your own user directory.
- Setting up "acquisition parameters" specifying that
the collection criteria must be set to "Number of Events
or Time", and the time must be set to less than 250 seconds.
The number of events can be any number, and depends upon the
frequency of your target population. This is automatically saved
with the CellQuest template.
Compensation control samples are set up in the usual way. If you
have stained all of your cells in a plate, you probably want to
transfer the compensation control samples to 12x75 FACS tubes before
running them manually in CellQuest. I had one technician figure
out how to run compensation samples from within MPM, but it's tricky
and I'm not sure I recommend it.
MPM Software
The MAS is controlled by the MPM software. As described
below, MPM works with CellQuest in a way that is
similar—but not
identical—to the way that Worklist Manager works
with CellQuest.
Setting the User
When you open the MPM software, you are asked to
select a user
from a pull-down menu. You can select "General User", or set
up a User for your self. The General User has access to the keyword
subsets
and protocols established by all users, but an individual user can
select from among only keyword subsets and protocols that were established
by that individual user.
Priming the system
Before running the MAS system, it is necessary to prime
the fluidics two-to-three times. You might as well do it at
the very beginning.
Setting initial and final washes
The initial and final washes flush the inside and outside of the
probe, CIU, and CIU tubing. You can set the MPM software to perform
0-4 washes. Use at least two. To set up washes, select the command:
Autosampler > Set
Initial and Final Washes
Plate configuration
The MAS system can acquire samples from a variety of different
plates, as noted in the Plate
Definition Guide. Since these plates
have different well depths, it is necessary to tell the MPM software
which plate you are using. Choose
Layout > Plate Configuration, and select your plate from the pull-down
menu.
Software settings
The following sections describe important software settings within
MPM.
Filenames
The filenames for your FCS files will all have
the same prefix (in contrast to filenames that
are used
in Worklist
Manager,
which have
prefixes that vary from sample to sample within
a panel). There are three fields in which you can
enter
parts
of the prefix.
I suggest
that you separate the parts of the prefix with
underscores ("_").
For part I of the prefix, I suggest that you use your initials, followed
by the date, in the following format: "JDA_020627_". The
second and third parts of the prefix can be used to indicate specific
information about the experiment, as in the following examples:
| Example Number |
Part II |
Part II Explanation |
Part III |
Part III Explanation |
| 1 |
RhM2_Gr1_ |
Rhesus experiment M2, Group 1 |
Week54 |
Week 54 of the experiment |
| 2 |
HuTetScreen_ |
Tetramer screen of a new donor |
|
This part might be blank |
| 3 |
LCMV_IFNg_ |
IFNg stimulation experiment in an LCMV-infected mouse |
Balbc |
This experiment was done in Balb/c mice. |
You don't need to use an underscore following the
last of the prefixes that you use; the MPM software
will
add a "." followed
by the well designation.
The filenames in MPM will sometimes contain less
information than filenames established by Worklist
Manager. For
example, if you
are analyzing
samples from many donors that were processed on
a single plate, it will not be possible to indicate
the donor
numbers in the
filenames. However, as we'll see below, this can
be done within keywords that
are contained within the FCS file.
Instrument Settings
Instrument settings files are established from
within CellQuest. In contrast to the situation
with Worklist
Manager, they
can be stored in any folder. I suggest that you
store them in a
folder that you
create
for yourself under "FACS Users".
CellQuest Template
Like Worklist Manager, MPM requires you to establish
a Template in CellQuest. The CellQuest Template
must be set
up to contain:
- Acquisition dot plots and histograms
- Any statistics
you wish to display during
acquisition
- Acquisition parameters.
In
the acquisition parameters panel, it
is necessary to
set to acquire "Number of Events or
Time", and the time should be set to 250 seconds
or less. However,
unlike the situation with Worklist Manager, the panel settings in
the Parameter Descriptions are not used by MPM. Consequently, the
labels
for you parameters are not displayed during acquisition, but they
are contained in the FCS files saved by MPM.
Data Storage Folder
It is necessary to tell MPM where to store your
FCS files. Do this by hitting the Data Storage
button
in the Setup
tab, then
navigate to the folder you would like to store
you data in.
Keywords: A Crucial Means for Annotating Your Data
Since the file names created by MPM for a single
experiment all contain the same prefix, it is more
crucial than
ever to use
Keywords to properly
annotate your data. You can use Keywords to indicate
properties such as donor numbers or names, mouse
strains, sample tissues,
staining
methods, time points in an experiment, etc. In
fact, the Keywords option in MPM is extremely flexible,
although how to set them
up and use them
is not obvious at the outset.
Defining a new Keyword
You can define a keyword such as "Sample Type: Human". You
can associate with each keyword a list of options that can be selected
from a pulldown menu. For example, members of the "Sample Type:
Human" list might include: (1) WB: EDTA, (2) WB:Heparin, (3) WB:ACD,
(4) PBMC: Fresh, (5) PBMC: Frozen, and (6) Lymph node. Another user
might have set up a different keyword called "Sample Type: Mouse" that
includes: (1) Splenocytes, (2) PBMC, (3) Thymocytes, and (4) Lymph
node. Either use might have set up a keyword such as "Staining
Protocol" that might include: (1) Surface, (2) ICC: Pharmingen,
and (3) ICC: Becton Dickinson.
Defining Keyword Subsets
In the paragraph above, it was obvious that users
performing experiments in mice and in humans might
wish to use
different keyword subsets.
These can be selected in....
Defining stain names
To be completed
Acquisition parameters used in MPM
Wash and mixing cycles
The wash cycle is used to minimize carryover from
one sample to another. The default value is for
a single
cycle only.
It remains
to be determined
if additional cycles are required.
The MAS unit will mix the samples in each well
by pipetting up and down. The default value is
for a
single cycle.
It remains to be determined
if additional cycles are required.
Sample volume
The Sample Volume parameter instructs the syringe
pump on the MAS to pull up the specified volume.
In order
to avoid
introducing
bubbles into the flow cell, this should normally
be at least 30 µl less
than the volume contained in the sample wells. For example, if your
sample wells contain 200 µl, then the largest value that you
should use for Sample Volume is 170 µl.
The maximum sample volume that can be used is 250 µl.
The dead volume of the system is approximately
30 µl. This means
that if your sample volume is set to 250 µl, you should expect
to acquire data for only 220 µl, or 12% of your sample. This
loss cannot be avoided.
Stuff below here to be edited
Steps for setting up and acquiring data.
17. In CellQuest, obtain file settings as usual.
18. In CellQuest, set up an "Acquisition Template". This
template contains plots, stats, and acquisition parameters. For the
acquisition parameters, the collection criteria must be set to "Number
of Events or Time", and the time must be set to less than 250
seconds. The number of events can be any number, and depends upon
the frequency of your target population.
19. The Instrument Settings file and the acquisition
template can be stored anywhere. I suggest that
you store them in
your own subfolder
under " FACS Users", rather than in a shared folder.
20. Quit CellQuest
21. Open the MPM software from the folder "BD Applications" in
the Apple Menu.
22. Select the "Acquisition" button from the first dialog
box.
23. Select yourself from the "User" pulldown menu in the
second dialog box, or click on the "New User" button if
this is you first time using MPM.
24. The MPM software takes some time to boot up.
Be patient
25. When the MPM window first appears, select the "AutoSampler
Parameters View" tab if is not selected.
26. Set the file name prefixes. You don't need
to use all three. I suggest that you use an underscore
after
all but
the last
of the prefixes.
The same prefixes will be used for all of your
samples.
27. Click the "Acquisition Doc" button and navigate through
the file system to find your acquisition template. Note that MPM
does not appear to use standard Macintosh dialog boxes.
28. Click the "Instr. Sett. File" button and navigate through
the file system to find your file settings.
29. Click in the "Sample Vol" field and enter a number that
is ≤ 250. Note that you should have at least 30 µl more
volume in your sample wells than the number you enter here.
30. Click in the "Mixing Vol" field, and enter a mixing
volume that is approximately 1/2 of the volume in your wells.
31. Click on the "Data Storage Folder" button at the upper
right and navigate through the file system till you decide where
you want to store your data. You can make new folders in this dialog
box.
32. From the "Current Plate" pull down menu, select you current
plate. The plate numbers are described in the document "Plate_Definition_Guide.pdf" that
is available on the Mordred or FACS-G4 computers. It is very important
that you use only those plates that are described in bold in this
document, and that you select the appropriate plate from the pull-down
menu.
This parameter determines the depth that is used by the sample probe.
33. Click on the "Keyword Subsets" tab.
34. To set up new keyword subsets, select "Modify > Modify
Master Keyword Set" from the "Layout" menu. This will
open a spreadsheet like dialog box. Click the "Add" button
to add new keyword fields. Click in the Keyword cell to change the
name of the keyword. Click on the "Edit Choices" button to
construct a list of keywords that can be selected from a pull-down
menu. Click on the "Choose From List" check box to force
choices from the menu.
35. When you have set up as many keyword fields
as are required for your experiment, select the
rows
that you
want to group
together in a "Keyword Subset". For example, you might have keywords
such as "Sample Type", "Donor Number", "Lysing
Protocol", and "Mouse Strain". For human experiments,
you might wish to group the first three together in a subset called "Human".
After selecting the rows, click on the "Define Keyword Subset" button,
and enter the name of the new subset in the "New Entry:" field.
Click on the OK button to leave this dialog, then on the "OK" button
again to leave the "Master Keywords Set" dialog box.
36. Select the cells that you'd like to describe
with a particular keyword. You can select the whole
plate
by clicking
in the
upper left hand corner of the plate, and you can
select rows or columns
by clicking
on the row or column headings. You can select single
cells by clicking in them, you multiple continguous
wells by
dragging across a number
of wells, or multiple discontinuous wells by holding
down the "Control" button
and clicking them (the programmer was obviously not a Mac person!).
37. For now, ignore the "Panels" and "Overlays" sections.
Panels doesn't work, and I don't know how to use "Overlays".
38. Click on the "Protocol Parameters View" tab. This is
where you'll enter you reagents for you fluorescent channels.
39. If all of your positions on your plate contain
different stains, these can be entered cell by
cell. Otherwise,
follow the instructions
below to reduce you typing.
40. Select the wells for the common reagents first.
For example, if you are using CD3 FITC and CD8
PerCP for
every sample,
click on the
corner of the plate layout to select all of the
wells, and then enter CD3 FITC in the FL1 field
and CD8
PerCP in the
FL3 field.
41. Next select wells for members of a panel. For
example if every sample in column 1 uses CD94 PE,
then select
column 1
and enter
CD94 PE. Do the same thing for the rest of the
columns, and then for the
rows.
42. You're almost ready to start your run. There
is one thing you need to do first, however.
43. PRIME THE SYSTEM! This is distinct from the "Prime" button
on the front of the cytometers. The Prime System command is available
from the "Autosampler" menu. Do this at least once, if
not twice, before you start your run.
44. Select the wells you'd like to run. This is
distinct from the keyword and protocol parameters
assignments
you've set
up. In addition,
you
can select the wells in any order, and they will
run in the order you've selected. Once you've select
the
wells,
do not
click on
the plate again.
45. Save your protocol. Select the "Save Protocol" command
from the Save menu. Protocols are stored in a "HD Name > BD
Applications > MPM > Users > UserName" folder.
46. Start your run. To do so, select "Acquire" from the Acquisition
menu. Then, click on the "Acquire" button.
Shut down
Run a 1/2 tube full of bleach for 5 minutes.
Run a 1/2 tube full of water for 5 minutes.
Shut down the computer, cytometers, and FACSFlow
supply system.
Trouble shooting
1. If you need to pause you run, I don't know how
the system handles this if done from within CellQuest
or
MPM.
2. I do know that you can hit "Pause" in MPM, and then
skip defined wells (for example, if you have a clog, etc.
Created on 6/9/02 12:56 PM 1 Modified: 8/31/03
|