|
Using
the Cary Spectrophotometer
This document describes the principles of using the Cary Spectrophotometer
at the Emory Vaccine Research Center and in the Emory Tetramer
Core.
Principles of spectrophotometry
Spectrophotometry is most often used in the biological research
laboratory to measure the concentration of some particular
substance. There is
only one equation that most of you will have to remember, and
it is called Beer's Law:

Where Ai is the absorbance (aka OD, or optical density) at wavelength "i",
is the "molar extinction coefficient" at wavelength "i",
and "l" is the pathlength of the "cell" or
cuvette. Since absorbance is a unitless quantity and concentration
is usually measured in "moles/liter" and
most cuvettes are usually 1 cm in length, the units of are
(liters / (mole*cm)).
The importance of absorbance maxima.
Although a substance will have an extinction coefficient
at every wavelength, concentrations are typically measured
at
maxima in
the absorbance spectra
because this is where the absorbance changes least
with changes in wavelength (i.e. the slope of the absorbance
spectra
is
zero).
Spectra versus individual wavelengths
In general, you can use simple wavelength
reads in cases such as determining where the peak of a protein coming
off of a chromatography column. If you want more information about
the quality of your substance, I recommend that you take
a full spectrum. When you do take a spectrum, be sure to instruct
the software on the spectrophotometer to pick peaks and report the
data at the peaks. I also recommend that you save these data files
on the spectrophotometer's computer.
The figure below shows a typical protein spectrum.
The commonly-found features include: (1) little-to-no absorbance
between 310-340 nm; (2) a shoulder at approximately 290 nm; (3)
a peak at approximately 280 nm; (4) a trough at approximately 245-250
nm; and (5) a rise between 245-240 nm. For a pure protein, the 260/280
ratio should be between 0.5-0.7.

Simple reads at specified wavelengths
If you wish to determine the concentration of a
substance in a number of fractions, you will
probably not collect
full spectra
for each fraction,
but instead you will collect data at a few
wavelengths, including the approximate absorption maximum.
Never, ever, attempt to measure the concentration
of a substance by collecting data at only
one wavelength (e.g. at 280 nm
for protein solutions). You should always
collect data
at multiple
wavelengths,
including one where you expect the absorbance
to be zero, and one that
will allow you to gain some idea about contaminating
substances. For proteins, you should collect
data at the following wavelengths:
1)
320 nm, where the absorbance of the protein
should be close to zero; 2) 280 nm, near
the absorption
maximum for most
proteins; and 3) 260
nm, which should give you some idea of the
level of contaminating
nucleic acids or other materials. Note that
these are the same wavelengths
that should be used to measure the concentrations
of DNA or RNA.Use
appropriate blanks
In most cases, you'll want to "zero" the spectrophotometer
using a cuvette filled with the same buffer as was used
for you sample solution. This is particularly necessary
if your
buffer contains
some light absorbing compounds such as DTT or Triton X-100.
However, many
buffers (PBS, Tris, HEPES) do not contain substances that
absorb light in the near UV range that we use most often,
and some leeway
is permitted
here. If you are attempting to measure the concentration
of a protein dissolved in a high concentration of a denaturant,
your blank should
also contain the same concentration of the denaturant.
Dynamic Range and Reliability of Spectrophotometric
Data
Although the spectrophotometer will never
fail to produce a number, it is essential
to remember
that
those numbers
are
not always
reliable. In particular, the spectrophotometer
will not give accurate and
meaningful results if the absorbances
are either too high or too low.
- Numbers below 0.020. These should
be regarded with suspicion, and
must be
supported with
sensible data
at additional wavelengths.
In
some cases, all you want is qualitative
information, so it won't matter
too much if you get accurate
quantitative results.
However,
where you
require accurate quantitative
data, you should attempt to either concentrate
your sample
or dilute it less.
- Numbers above 1.5. The spectrophotometer
cannot accurately measure absorbances
much above 1.5.
If the absorbance
of your sample is
above 1.5, you should further
dilute you sample and measure the absorbance
on it again.
Common artifacts
The following are the most common
causes of artifacts in spectrophotometric
measurements:
- Air bubbles in the cuvette.
Microcuvettes are particularly
susceptible to collection
of air bubbles
in the sample
beam. You should always
make sure that there
are no air bubbles in the sample beam.
- Particulates in the sample.
Particulate matter
in the sample will scatter
light and result in
absorbance readings
that are
significantly higher
than the true reading. Note
that particulates are particularly
likely to arise if
you attempt to dilute denaturant (urea,
guanidinium chloride)
solubilized proteins
into solutions
not containing
denaturants; this should
be
avoided, as noted below.
- Contaminating substances
in diluents. Be very
careful about
the presence
of light absorbing
substances in
the solutions you
use to dilute the substances that
you wish to measures.
One substance that is
often
forgotten is
DTT, which will
absorbe UV light
when oxidized
to the disulfide
form.
- Incomplete mixing
of samples can yield
improper
results.
Although this seems
like a no-brainer,
it can be a matter
of serious concern
when you are preparing
sample dilutions
with small volumes
(e.g. 2 µl
into 98µl).
- Dirty cuvettes.
The cuvettes
should be clean
on both the
outside and
the inside. The
laboratory should
have
a cuvette
washer,
and cuvettes
should be washed and then
rinsed
with clean, high
quality solvents.
- Residual liquid
in the cuvette
before addition
of the sample.
This will cause
either a) dilution
of the
sample,
resulting
in an artifactually
low reading;
or b) contamination of
the sample with
a solution
containing a "high absorbing" compound,
resulting in
an artifactually
high reading.
Measuring the concentration
of denaturant
solubilized proteins
Dilute denaturant
solubilized
proteins only into
denaturants.
Otherwise,
your samples
will likely
precipitate
and you'll end up
measuring light
scatter rather
than the absorbance
of
your
sample. Urea
solutions, in particular, should be freshly prepared.
Measuring absorbance
of fluorescent
compounds,
including
fluorophore-labeled antibodies
Know the
absorbance
of you
fluorophore, which
can be
found somewhere
in the
literature or on the
web. The
Molecular
Probes
web site (http://www.probes.com)
contains
fluorescence
spectra
for a very large
number
of compounds).
Protocol:
Using
the Cary
Spectrophotometer
To be filled in.
|